CORRESPONDENCE| Volume 55, ISSUE 3, P400-402, April 2023

Download started.

Ok

Interference of methylene blue therapy on methaemoglobin measurement

Published:November 07, 2022DOI:https://doi.org/10.1016/j.pathol.2022.07.018
      To the Editor,
      Methylene blue (MB) is a classical antidote for treating methaemoglobinaemia. Conventional co-oximetry utilises multi-wavelength spectrophotometer for methaemoglobin (MetHb) determination, which shares a similar absorption spectrum to MB.
      • Cruz-Landeira A.
      • Bal M.J.
      • Quintela
      • López-Rivadulla M.
      Determination of methemoglobin and total hemoglobin in toxicological studies by derivative spectrophotometry.
      Previous studies demonstrated significant analytical interference of MB on MetHb measurement, but the studies were conducted more than 20 years ago, and in one study the MB concentrations investigated were at supra-therapeutic range.
      • Gourlain H.
      • Buneaux F.
      • Borron S.W.
      • Gouget B.
      • Levillain P.
      Interference of methylene blue with co-oximetry of hemoglobin derivatives.
      ,
      • Dötsch J.
      • Demirakça S.
      • Hahn D.
      • Katz N.
      • Kühl P.G.
      • Rascher W.
      Accuracy of methemoglobin measurements: comparison of six different commercial devices and one manual method.
      With the emergence of newer co-oximetry models and better pharmacokinetic profiling of MB, this study aims to re-investigate the effect of MB interference on MetHb measurement, by spiking clinically relevant MB concentrations to heparinised whole blood samples at varying MetHb concentrations.
      In the current study, MetHb concentration in heparinised whole blood sample was measured with ABL800 FLEX co-oximetry (Radiometer, Denmark), which utilised a 128-wavelength spectrophotometer, at a measurement range of 478–672 nm.
      Radiometer Medical ApS. ABL800 FLEX reference manual. Denmark: Radiometer Medical ApS
      Optical measuring principles, performance characteristics.
      The analytical coefficient of variation (CVA) of the two co-oximetry analysers used in this study was estimated to be 1.6–2.1% at MetHb concentrations of 11.5–28.5% unit.
      MetHb samples were prepared in-house by treating heparinised whole blood samples with sodium nitrite solution as described by Shihana et al.
      • Shihana F.
      • Dissanayake D.M.
      • Buckley N.A.
      • Dawson A.H.
      A simple quantitative bedside test to determine methemoglobin.
      and Lim et al.
      • Lim S.F.
      • Tan I.K.
      Quantitative determination of methaemoglobin and carboxyhaemoglobin by co-oximetry, and effect of anticoagulants.
      In brief, sodium nitrite solutions were prepared by dissolving sodium nitrite standard (BDH, Prod 102564W, Lot A608266 617; VWR International Ltd, USA) in normal saline (0.9% sodium chloride) to a final concentration of 100, 500 and 2000 mg/L. Two hundred μL of one of these sodium nitrite solutions was added to 1 mL of heparinised whole blood sample collected from healthy volunteers. The spiked blood samples were mixed and incubated for 5–10 min at room temperature before subjecting to MetHb measurement. Blood samples with MetHb concentrations from approximately 10%–60% unit were prepared.
      MB stock solution of 10 mg/mL (DIN:02230770, Lot L0010072; Omega, Canada) was diluted with normal saline to obtain working solutions at final concentrations of 10, 100 and 400 mg/L. Sixty μL of one of these working MB solutions was added to 1.2 mL in-house prepared MetHb blood samples to reach final MB concentrations of 0.5, 5 and 20 mg/L. Control aliquots were prepared for each data point, by spiking 60 μL normal saline without MB. Each sample was compared with its own control and MetHb measurement was performed immediately after treatment of the blood sample with the spike. The absolute difference (% unit) and relative difference (%) in MetHb measurement between the MB-spiked sample and its corresponding control were calculated as below.
      Absolute difference in MetHb concentration (% unit) = MetHb concentration in MB-spiked sample (% unit) – MetHb concentration in corresponding control sample (% unit).
      Relative difference in MetHb concentration (%) = Absolute difference in MetHb concentration (% unit) ÷ MetHb concentration in control sample (% unit) × 100%.
      With reference to the Analytical Performance Specifications of the Royal College of Pathologists of Australasia Quality Assurance Programs and our in-house CVA on MetHb measurement, we define a significant difference in MetHb concentrations measured between the spiked and control samples as ±2.0% unit (absolute difference) at MetHb concentrations less than or equal to 20% unit, and ±10% (relative difference) at MetHb concentrations greater than 20% unit.
      • Royal College of Pathologists of Australasia Quality Assurance Programs
      Chemical pathology analytical performance specifications.
      Results for MetHb concentrations measured in the MB spiked samples and the corresponding control samples were listed in Table 1.
      Table 1MetHb concentrations measured in blood samples with or without MB spiking
      MetHb concentration(s) measured (% unit)Absolute difference(% unit)
      Absolute difference in MetHb concentration (% unit) = MetHb concentration in MB-spiked sample (% unit) – MetHb concentration in corresponding control sample (% unit).
      Relative difference (%)
      Relative difference in MetHb concentration (%) = Absolute difference in MetHb concentration (% unit) ÷ MetHb concentration in control sample (% unit) × 100%.
      Sample with MB spikedControl sample with no MB spiking
      MB spiked at 0.5 mg/L
       MetHb level 113.913.7+0.2+1.5
       MetHb level 236.336.0+0.3+0.8
       MetHb level 346.146.6–0.5–1.1
       MetHb level 461.963.3–1.4–2.2
      MB spiked at 5 mg/L
       MetHb level 110.714.7–4.0–27.2
       MetHb level 237.338.7–1.4–3.6
       MetHb level 339.539.3+0.2+0.5
       MetHb level 450.452.9–2.5–4.7
      MB spiked at 20 mg/L
       MetHb level 1Interfere
      Error tag was shown on this measurement, and the MetHb concentration in this sample could not be measured.
      13.3
       MetHb level 219.333.5–14.2–42.4
       MetHb level 353.453.1+0.3+0.6
       MetHb level 465.366.8–1.5–2.2
      a Absolute difference in MetHb concentration (% unit) = MetHb concentration in MB-spiked sample (% unit) – MetHb concentration in corresponding control sample (% unit).
      b Relative difference in MetHb concentration (%) = Absolute difference in MetHb concentration (% unit) ÷ MetHb concentration in control sample (% unit) × 100%.
      c Error tag was shown on this measurement, and the MetHb concentration in this sample could not be measured.
      Previous studies by Gourlain et al.
      • Gourlain H.
      • Buneaux F.
      • Borron S.W.
      • Gouget B.
      • Levillain P.
      Interference of methylene blue with co-oximetry of hemoglobin derivatives.
      and Dötsch et al.,
      • Dötsch J.
      • Demirakça S.
      • Hahn D.
      • Katz N.
      • Kühl P.G.
      • Rascher W.
      Accuracy of methemoglobin measurements: comparison of six different commercial devices and one manual method.
      published before the millennium, demonstrated that MB posed a significant negative interference on MetHb measurement with a conventional ABL co-oximetry (ABL520 model, utilising 6 wavelengths for measurement). In Gourlain's study, the effect of spiking MB, at concentrations of 10–100 mg/L, into blood samples with MetHb concentrations of 4.6–14.5% unit was evaluated. The MetHb concentrations measured in the MB-spiked samples showed an absolute difference of –3.1% to –34% unit, when compared to the corresponding unspiked samples. However, these data may not be readily applicable to our current laboratory practice, as the modern co-oximetry models utilise more wavelengths for measurement and are theoretically less susceptible to spectroscopic interference by MB (ABL800 FLEX, utilising 128 wavelengths for measurement). Moreover, with improved understanding of MB pharmacokinetic profiling, MB spiked in the above study were actually at supra-therapeutic concentrations.
      • Peter C.
      • Hongwan D.
      • Küpfer A.
      • Lauterburg B.H.
      Pharmacokinetics and organ distribution of intravenous and oral methylene blue.
      Furthermore, the manufacturer manual of the co-oximetry used in the current study only provided limited data concerning the potential interference of MB on MetHb measurement. According to the manufacturer's claim based on mathematical superimposition of interference spectra and calculations, MB at 30 mg/L was predicted to cause a negative interference in MetHb measurement with an absolute difference of –6.2% unit in a sample with 0.5% unit MetHb.
      Radiometer Medical ApS. ABL800 FLEX reference manual. Denmark: Radiometer Medical ApS
      Optical measuring principles, performance characteristics.
      Still these data are not clinically useful, as MB interference was only estimated at a single supra-therapeutic concentration at which the MetHb level was clinically insignificant.
      In this study, the magnitude of MB interference at therapeutic concentrations on MetHb measurement was investigated. The conventional MB dosage for treating methaemoglobinaemia was 1–2 mg/kg given intravenously. Pharmacokinetic data showed that whole blood MB level reached a peak concentration of 3.2 mg/L, 5 min after intravenous administration of 100 mg MB, and eventually dropped to 0.3 mg/L at 30 min post-administration.
      • Peter C.
      • Hongwan D.
      • Küpfer A.
      • Lauterburg B.H.
      Pharmacokinetics and organ distribution of intravenous and oral methylene blue.
      Therefore, MB concentrations of 0.5 mg/L and 5 mg/L were chosen for this study to mimic the blood concentrations after a therapeutic dose of MB. The study was also performed at a higher concentration of MB (20 mg/L) to replicate findings from previous interference studies by Gourlain et al.
      • Gourlain H.
      • Buneaux F.
      • Borron S.W.
      • Gouget B.
      • Levillain P.
      Interference of methylene blue with co-oximetry of hemoglobin derivatives.
      and Dötsch et al.
      • Dötsch J.
      • Demirakça S.
      • Hahn D.
      • Katz N.
      • Kühl P.G.
      • Rascher W.
      Accuracy of methemoglobin measurements: comparison of six different commercial devices and one manual method.
      The concentrations of MetHb chosen in this study were based on the relationship between MetHb concentrations, clinical manifestations and treatment threshold. Patients typically start to present with cyanosis at MetHb concentration of 10–20% unit; MB treatment is conventionally started in symptomatic patients with MetHb concentration >20% unit, or asymptomatic patients with MetHb >30% unit; and MetHb concentration >70% unit is usually fatal.
      • Khanal R.
      • Karmacharya P.
      • Pathak R.
      • Poudel D.M.
      • Ghimire S.
      • Alweis R.
      Do all patients with acquired methemoglobinemia need treatment? A lesson learnt.
      Therefore, we evaluated the effect of MB on samples with MetHb concentrations at 10–60% unit, which was the clinically significant range. In our study, MetHb measurement was made immediately after the addition of MB. This helped minimise the time for chemical reduction of MetHb, thereby allowing the investigation of pure physical interferences. The calculated difference in MetHb concentration between the sample and control set-up therefore was attributed solely to the spectroscopic interference present.
      At MB concentration of 0.5 mg/L, mimicking the typical therapeutic MB concentration 30 min after antidote administration, no significant interference was observed. Given the onset time for MB's action of 30 min,
      • Su Y.F.
      • Lu L.H.
      • Hsu T.H.
      • Chang S.L.
      • Lin R.T.
      Successful treatment of methemoglobinemia in an elderly couple with severe cyanosis: two case reports.
      it could be safely concluded that the interference on MetHb measurement would be minimal if blood taking is arranged 30 min after giving the MB antidote. In clinical practice, after MB treatment, patients with methaemoglobinaemia are monitored for resolution of symptoms and with serial MetHb measurements.
      • Ash-Bernal R.
      • Wise R.
      • Wright S.M.
      Acquired methemoglobinemia: a retrospective series of 138 cases at 2 teaching hospitals.
      Repeated blood testing for MetHb is usually made in 1–3 h.
      • Su Y.F.
      • Lu L.H.
      • Hsu T.H.
      • Chang S.L.
      • Lin R.T.
      Successful treatment of methemoglobinemia in an elderly couple with severe cyanosis: two case reports.
      ,
      • Caudill L.
      • Walbridge J.
      • Kuhn G.
      Methemoglobinemia as a cause of coma.
      The results from our study are therefore in-line with the current practice, and could help reassure clinicians on the accuracy of MetHb concentration measured.
      At MB concentration of 5 mg/L, which simulated the maximal MB concentration at 5 min after antidote administration, MetHb concentrations at >20% (levels 2, 3 and 4) were relatively unaffected. In contrast, a significant negative interference (absolute difference of –4.0% unit, equivalent to 27.2% relative decrease from the control aliquot level) was observed when MetHb concentration was <20% unit (level 1). Therefore, clinicians should be advised not to recheck MetHb concentration immediately after MB administration, as this may misguide clinical judgement and any subsequent management plan.
      At MB concentration of 20 mg/L, a significant negative interference (absolute difference of –14.2% unit, equivalent to 42.4% relative decrease from the control aliquot level) was observed at lower MetHb concentration at approximately 30% unit (levels 2). At level 1 with MetHb concentration of approximately 10%, MB spiking even introduced an error signal and the MetHb level in this spiked sample could not be determined. Higher MetHb concentrations were relatively free from MB interference (levels 3 and 4). These data were consistent with the findings by Gourlain et al.
      • Gourlain H.
      • Buneaux F.
      • Borron S.W.
      • Gouget B.
      • Levillain P.
      Interference of methylene blue with co-oximetry of hemoglobin derivatives.
      and Dötsch et al.
      • Dötsch J.
      • Demirakça S.
      • Hahn D.
      • Katz N.
      • Kühl P.G.
      • Rascher W.
      Accuracy of methemoglobin measurements: comparison of six different commercial devices and one manual method.
      Modern co-oximetry, despite the improved technical specifications, is still subjected to MB interference at a very high concentration. However, this phenomenon is unlikely clinically significant as such MB concentration is much higher than the usual therapeutic range.
      Several limitations existed in the current study. First, our interference study was carried out on the ABL800 FLEX co-oximetry only. It may not be possible to generalise our findings to other analysers. For laboratories using other models of co-oximetry it may be worth performing independent studies to evaluate the susceptibility of their analysers to MB interference. Second, the MetHb concentrations in the blood samples were not equally spaced, and they differed slightly between spiking set-ups. This could be attributed to the minor variations in incubation time and temperature. Third, in order to perform simultaneous MetHb measurement on both the spiked and its corresponding control samples, two independent blood gas analysers were employed, which may have introduced a larger CVA.
      In conclusion, an interference study of varyingly spiked MB concentrations on measurement of MetHb at 10–60% unit using ABL800 FLEX was performed. No interference was observed on MetHb measurement at the usual therapeutic concentration of 0.5 mg/L MB. A significant negative interference could be observed on MetHb measurement when MB was spiked at 5 mg/L, simulating the peak concentration after an intravenous dose. Clinicians are advised not to recheck MetHb concentration immediately after MB administration. However, they can be reassured that the measured MetHb concentration is not subjected to any significant MB interference 30 min after intravenous MB administration.

      Conflicts of interest and sources of funding

      The authors state that there are no conflicts of interest to disclose.

      References

        • Cruz-Landeira A.
        • Bal M.J.
        • Quintela
        • López-Rivadulla M.
        Determination of methemoglobin and total hemoglobin in toxicological studies by derivative spectrophotometry.
        J Anal Toxicol. 2002; 26: 67-72
        • Gourlain H.
        • Buneaux F.
        • Borron S.W.
        • Gouget B.
        • Levillain P.
        Interference of methylene blue with co-oximetry of hemoglobin derivatives.
        Clin Chem. 1997; 43: 1078-1080
        • Dötsch J.
        • Demirakça S.
        • Hahn D.
        • Katz N.
        • Kühl P.G.
        • Rascher W.
        Accuracy of methemoglobin measurements: comparison of six different commercial devices and one manual method.
        Crit Care Med. 1999; 27: 1191-1194
        • Radiometer Medical ApS. ABL800 FLEX reference manual. Denmark: Radiometer Medical ApS
        Optical measuring principles, performance characteristics.
        2012
        • Shihana F.
        • Dissanayake D.M.
        • Buckley N.A.
        • Dawson A.H.
        A simple quantitative bedside test to determine methemoglobin.
        Ann Emerg Med. 2010; 55: 184-189
        • Lim S.F.
        • Tan I.K.
        Quantitative determination of methaemoglobin and carboxyhaemoglobin by co-oximetry, and effect of anticoagulants.
        Ann Clin Biochem. 1999; 36: 774-776
        • Royal College of Pathologists of Australasia Quality Assurance Programs
        Chemical pathology analytical performance specifications.
        2021 (cited Mar 2022)
        • Peter C.
        • Hongwan D.
        • Küpfer A.
        • Lauterburg B.H.
        Pharmacokinetics and organ distribution of intravenous and oral methylene blue.
        Eur J Clin Pharmacol. 2000; 56: 247-250
        • Khanal R.
        • Karmacharya P.
        • Pathak R.
        • Poudel D.M.
        • Ghimire S.
        • Alweis R.
        Do all patients with acquired methemoglobinemia need treatment? A lesson learnt.
        J Community Hosp Intern Med Perspect. 2015; 529079
        • Su Y.F.
        • Lu L.H.
        • Hsu T.H.
        • Chang S.L.
        • Lin R.T.
        Successful treatment of methemoglobinemia in an elderly couple with severe cyanosis: two case reports.
        J Med Case Rep. 2012; 6: 290
        • Ash-Bernal R.
        • Wise R.
        • Wright S.M.
        Acquired methemoglobinemia: a retrospective series of 138 cases at 2 teaching hospitals.
        Medicine (Baltimore). 2004; 83: 265-273
        • Caudill L.
        • Walbridge J.
        • Kuhn G.
        Methemoglobinemia as a cause of coma.
        Ann Emerg Med. 1990; 19: 677-679